Negative Reference Voltage; Temperature Sensor - Emerson Rosemount Xstream Series Instruction Manual

Gas analyzer
Hide thumbs Also See for Rosemount Xstream Series:
Table of Contents

Advertisement

X-STREAM

8-4-3-1-3 Negative reference voltage

Measuring Point:
Measuring instrument:
Signal:
inverse positive
reference voltage
Failure:
Both reference voltages differ by
more than 10 mV
(U
ref. pos.

8-4-3-1-4 Temperature sensor

Measuring Point:
Measuring instrument:
Signal:
approx. 0 ± 500 mV DC
(at ambient temperature)
Failure:
Signal not within stated range
Potential cause:
-
IR measurement or paramagnetic
Oxygen measurement
a)
sensor not connected
b)
sensor defective
c)
sensor cable broken
d)
BKS board defective
-
electrochemical Oxygen measurement:
a)
sensor not connected
b)
OXS board defective
c)
BKS board defective
8-14
8-4
Troubleshooting on Components
X5, pin 4
DVM
+ U
< 10 mV) !
ref. neg.
X8
DVM
How to solve:
Replace BKS board
How to solve:
a)
connect temperature sensor
b)
replace temperature sensor
c)
replace temperature sensor
d)
replace BKS board
a)
connect temperature sensor
b)
replace OXS board
c)
replace BKS board