Preparation of signal cables
All signal cables are to be connected via screw
terminals, located at the analyzer's rear panel.
The strain reliefs provide a metal strip to connect
Supported wire cross sections:
Cable skinning length:
Hole diameter:
Screw thread:
Tightening torque, min:
Signal inputs / outputs
Analog signal outputs
To connect the terminals remove the cover at
the analyzer's rear panel (4 screws). Feed the
analog signals cable through the uppermost
edge protection, and through the upper strain-
relief.
The upper 4 terminals (# 1 - 4) of the terminals
row next to the power terminals are reserved
for analog signal outputs.
4-14
4-4-1 Installation - X-STREAM GP, X-STREAM GPS
(+) 4 (0) - 20 mA, channel 1
(-) 4 (0) - 20 mA, channel 1
(+) 4 (0) - 20 mA, channel 2
(-) 4 (0) - 20 mA, channel 2

Fig. 4-11: Analog Signal Output Terminals

to the cable shield after the cable's outer
insulation is stripped.
24 to 14 AWG (0.2 to 2.5 mm
no need to use wire end sleeves
0.354 inch (9 mm)
0.05 inch (1.2 mm)
M 2.5
3.5 in.lb (0.4 Nm)
Note!
Take care of the special installation
instructions in section 4-5!
Interface specification:
5-4-3-4-1 Analog output setup, page 5-31
≤ 500 Ω
Burden: R
B
Instruction Manual
Emerson Process Management GmbH & Co. OHG
HASAxE-IM-HS
07/2006