X-STREAM
Signal:
SignalRange:
Scaling..
Trim output..
Option
C1-C2
C1-T
T-C2
20 mA
0 / 4 mA A signal is applied to both analog outputs
*) Select one of these modes only to perform a quick check if the analog
output level settings are still correct.
6-8
6-2 Checking the Instrument's Setup
C1-C2
4-20mA
Output 1 source
Channel 1 gas
concentration
Channel 1 gas
concentration
Temperature sensor
"Temp-1" value
A signal is applied to both analog outputs
creating a 20 mA output signal
creating an output signal of 0 mA (when in
Dead-Zero mode) or 4 mA when in
Live-Zero mode.
Dead-Zero or Live-Zero mode is
selected by the "Signal range" parameter
(
page 6-8)
Select the "Analog output" line to open a
submenu to specify the analog outputs: The two
analog outputs may be configured to output
several signals via the "Signal" entry:
Output 2 source
Channel 2 gas
concentration
Temperature sensor
"Temp-1" value
Channel 2 gas
concentration
The next line "Signal Range" allows to configure
the analog outputs ranges / the analog outputs
to comply with the NAMUR NE43 recom-
mendations as described on the next page.
Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASAxE-IM-HS
05/2006
Operational
Modes
Modes for
checking the
output signal
adjustments
*)
only