5-4-3-1-3 Component Setup Menu

Setup..
Display..
Component..
Component
Tag:
DecimalPlaces:
CO2.1:
134.1 ppm
1
st
menu page
Unit
Text:
Factor:
Offset:
nd
2
menu page
Note!
Tag and unit text strings as well as factor and
offset values are not checked for plausibility!
The user may enter whatever data he wants!
Emerson Process Management GmbH & Co. OHG
5-4 Menu System - Display Setup Menu
CO2.1
1
ppm
1.0000
0.0000
Dual channel instrument:
Optional gas component selection menu -
Select the component to be setup
Enter the tag to be associated with the gas
component: Each digit has to be selected and
set independently!
Note!
".1" in this example is used to show that CO
is the first measuring channel (maybe helpful
for dual channel instruments, but not required).
Set measuring value's decimal places:
0 to 4
Current sample gas concentration taking into
account the settings above. Format is updated
the moment setup changes are entered.
Enter the text string to be associated with the
gas component concentration unit: Each digit
has to be selected and set independently!
Internally the analyzer handlesconcentrations in
ppm. For showing other units enter the factor
to convert ppm into the required unit, e.g.
0.0001 for showing %.
If required, set an offset to be added to the
measured value.
Dual channel instrument:
Pressing the
key while any line is selected
returns to the optional gas component selec-
tion menu to open the same menu for the
other measuring channel.
X-STREAM
2
5-25