Defining a window
Defining a window between the high and the
low threshold level (fig. 6-1) gives an alarm when
the current concentration either exceeds the
upper limit (area "D") or falls below the lower
limit (area "B").
Maximum one alarm is activated at a time for
each channel!
Standard mode:
If an alarm is activated, the corresponding
alarm outputs a "High" level signal.
Settings:
Level 1 > level 2
Level 1 function: High
Level 2 function: Low
Fail-safe mode:
If an alarm is activated, the corresponding
alarm outputs a "Low" level signal.
Settings:
Level 1 > level 2
Level 1 function: High FS
Level 2 function: Low FS
As long as any alarm is activated a status
message appears in the measuring screen's
th
4
line.
Emerson Process Management GmbH & Co. OHG
6-2 Checking the Instrument's Setup

Fig. 6-1: Thresholds defining a window