Analog Preamplifier - Emerson Rosemount Xstream Series Instruction Manual

Gas analyzer
Hide thumbs Also See for Rosemount Xstream Series:
Table of Contents

Advertisement

X-STREAM

8-4-3-1-6 Analog Preamplifier

a)
Paramagnetic Oxygen Measurement
Measuring Point:
Measuring instrument:
Signal when connected to
zero gas:
ambient air (approx. 21 Vol. - % O
100 % O2 sensor:
25 % O2 sensor:
(Type of sensor: see separate nameplate label)
Failure: No signal or wrong measuring values
Potential cause:
a)
Oxygen sensor not connected
b)
Oxygen sensor defective
c)
BKS board defective
b)
IR Measurement
Measuring Point:
X 25 (channel 1; not for instruments with
Oxygen measurement)
X 27 (channel 2)
Measuring instrument:
Signal when connected to
zero gas: 0 V dc (± 100 mV)
Zero point voltage and span voltage must differ
at least by 600 mV (at ranges < 1000 ppm
difference should be 500 mV min.)
Failure: No signal or wrong measuring values
Potential cause:
a)
Detector not connected
b)
Detector defective
c)
BKS board defective
8-16
8-4
Troubleshooting on Components
X 25 (channel 1)
DVM
0 V dc (± 50 mV)
):
2
approx. 840 mV
approx. 3.36 V
DVM
How to solve:
a)
Connect Oxygen sensor
b)
Replace Oxygen sensor
c)
Replace BKS board
How to solve:
a)
Connect detector
b)
Replace detector
c)
Replace BKS board
Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASAxE-IM-HS
05/2006