X-STREAM
Protocol:
MODB mode:
ID number:
Interface: RS485/2w
Baud rate:
Parity:
6-2-5
Concentration Alarms Setup
Level1:
Function:
Level2:
Function:
6-12
6-2 Checking the Instrument's Setup
MODB RTU
32Bit
2
19200
None
100 ppm
Low
500 ppm
High
Press the LEFT key several times to return to the
SETUP MENU, select the "Communication.." line
and enter the menu.
Note!
If you do not want to make use of concentration
alarms, skip the next pages and continue with
page 6-18.
Press the LEFT key several times to return to
the SETUP MENU, select the "Alarms.." line,
enter the menu and select the channel to be
setup (multi-channel instruments only).
Two different concentration thresholds can be
set for each channel, freely configurable within
the limits given by the "SpanRange" parameter.
The moment the measured concentration
exceeds one of the thresholds, a status message
appears in the measuring screen's 4
the related digital output (option) is set.
Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASAxE-IM-HS
05/2006
th
line and