5-4-3-3-3 AD Conversion Setup Menu

Setup..
Measurement..
AD conversion..
AD conversion
Comp1Offset
Comp2Offset
MuxOffset
5-32
If system is setup accordingly access level 3
code must be entered to access this menu.
Select the first line to setup the AD converters
offset for the first channel: The offset's unit are
counts.
Select the second line to setup the AD
20
converters offset for the second channel: The
20
offset's unit are counts.
20
Select the last line to setup the AD converters
offset for the secondary measurement (This is
a multiplexed converter): The offset's unit are
counts.
Instruction Manual
Changing parameters shown in
this screen is only required
when the analyzer's main board
BKS has been replaced: The
built-in AD converters have
differing unit-to-unit offsets,
therefore the data of the cur-
rently used units have to be
entered.
The offsets are given on a label
placed on the main board BKS.
Entering wrong values causes
wrong concentration measure-
ment results!
Replacing the main board is
permitted to trained personnel
only!
Emerson Process Management GmbH & Co. OHG
HASAxE-IM-HS
05/2006