X-STREAM
Modbus-Register
Daniel /
Enron
8001..8012
8021..8032
3092
3093
3094
3106
9023
9024
3070
4006
9063..9066
9067..9070
3090
9-10
9-3
Modbus - List of Parameters and Registers
Modicon
MeasBufValues#DB
5001..5024
5041..5064
ring buffer which contains single raw
measurement values
3092
Signal assigned to Line1 of measurement
3093
Signal assigned to Line2 of measurement
3094
Signal assigned to Line3 of measurement
MotorTimerDB
3106
defines the turning frequency of the chopper
NormConcentration#DB
6045..6046
6047..6048
normalized concentration value
NumberChannelsDB
3070
how many channels are built-in
ParamAccessModeDB
4006
parameter write access mode the analyzer is
PfactCorrCoeffs#DB
6125..6132
polynom coefficients that compute span factor
6133..6140
correction (PressureFactor#DB) using the
assigned pressure
(PressureSensorInstalledDB)
PresDecimalPointDB
3090
decimal point position for pressure displays
Mnemonic
Description
MeasLine1DB
display
MeasLine2DB
display
MeasLine3DB
display
motor
currently in
Instruction Manual
Range / Enum values / Coefficients
0 = NoSignal
1 = Comp1
2 = Temp1
3 = Press1
4 = Comp2
5 = Temp2
6 = Press2
0 = NoSignal
1 = Comp1
2 = Temp1
3 = Press1
4 = Comp2
5 = Temp2
6 = Press2
0 = NoSignal
1 = Comp1
2 = Temp1
3 = Press1
4 = Comp2
5 = Temp2
6 = Press2
1..2
0=normal
1=local access
2 = exclusive remote access
4=service access
A0, A1, A2, A3
0..4
Emerson Process Management GmbH & Co. OHG
HASAxE-IM-HS
05/2006
1
Attribute
Type
Dyn
Lword-Arr12
Cfg
Enum
Cfg
Enum
Cfg
Enum
Cfg, Fac
Word
Dyn
Float
Cfg, Fac
uchar
Dyn
Enum
Cfg, Fac
Float-Arr4
Cfg
uchar