Emerson Rosemount Xstream Series Instruction Manual page 263

Gas analyzer
Hide thumbs Also See for Rosemount Xstream Series:
Table of Contents

Advertisement

Instruction Manual
HASAxE-IM-HS
05/2006
Modbus-Register
Daniel /
Modicon
Enron
3078
3088
1016
3056
3059
3068
3069
3126..3129
3126..3129
3130
9033
6065..6066
9034
6067..6068
1013
3116..3119
3116..3119
3120
3055
1018
3001
3060
Emerson Process Management GmbH & Co. OHG
9-3
Modbus - List of Parameters and Registers
Mnemonic
Description
DecimalPoint#DB
3078
3088
decimal point for concentration display
DeviceStateDB
1016
measurement is running without errors
DifferentialMeasurementDB
3056
min/max values for differential measurement is
used
DigInputsInstalledDB
3059
The optional digital inputs are installed
DynamicNoiseReduction#DB
3068
3069
value for dynamic noise reduction (0 = Off)
EmersonAccessCodeDB
user code for getting access to Emerson areas
EmersonAccModeDB
3130
mode for getting access to Emerson areas
EndOfRange#DB
end of measurement range in ppm
EpromErrorDB
1013
checksum of EPROM is erroneous
ExpertAccessCodeDB
user code for getting access to expert areas
ExpertAccModeDB
3120
mode for getting access to expert areas
FlowAlarmInstalledDB
3055
flow alarm hardware installed
FlowTooLowDB
1018
too low flow is indicated
FlushingPeriodDB
3001
purge delay time (in secs) for gas supply
HideOptionLinesDB
3060
defines if menu lines are hidden when
depending on installed options
X-STREAM
Range / Enum values / Coefficients
0 = no digits after DP
1 = 1digit after DP
2 = 2digits after DP
3 = 3digits after DP
4 = 4digits after DP
bit0 = differential measurement Ch1
bit1 = differential measurement Ch2
0 = not installed
1 = installed
0 .. 9999
2 characters per register:
st
1
character at HIGH byte,
nd
2
character at LOW byte.
0 = access is allowed
1 = access requires user code
2 = access is prohibited
1 .. 1000000
2 characters per register:
st
1
character at HIGH byte,
nd
2
character at LOW byte.
0 = access is allowed
1 = access requires user code
2 = access is prohibited
0 = not installed
1 = installed
0..600
0..1
1
Attribute
Type
Cfg
Enum
Dyn
Boolean
Cfg, Fac
Byte-Bitfield
Cfg, Fac
Boolean
Cfg, Fac
Integer
Cfg, Fac
String-8
Cfg, Fac
enum
Cfg, Fac
Float
Dyn
Boolean
Cfg
String-8
Cfg
enum
Cfg, Fac
Boolean
Dyn
Boolean
Cfg
Integer
Cfg
Boolean
9-7

Advertisement

Table of Contents
loading

This manual is also suitable for:

X-stream fX-stream gpsX-stream gp

Table of Contents