Instruction Manual
HASAxE-IM-HS
07/2006
Relay status Signals
Design: dry relay contacts
Electrical specification: max. 30 V
Note!
Take care of the special installation instruc-
tions in section 4-5 and of the cable gland
assembly instruction on page 4-22!
For accessing the terminals open the
instrument's front door.
The middle 6 terminals (# 5 - 10) of the
terminals row next to the power terminals are
reserved for the relay status signals.
Terminal Signal
5
Failure COM
6
Failure NO
7
Maintenance, off-spec COM
8
Maintenance, off-spec NO
9
Function check COM
10
Function check NO
Emerson Process Management GmbH & Co. OHG
4-4-2 Installation - X-STREAM F
, 1 A, 30 W
Fig. 4-22: X-STREAM F - Relay Status Terminals
Fig. 4-21: Relay status signals, block diagram
10
9
8
7
6
X-STREAM
Signal output (NC);
configuration on
request
Signal output (NO);
factory standard
configuration
5
4-25