5-4-3-9 Miscellaneous Screen

Setup..
Miscellaneous..
Hold on cal:
5-46
5-4 Menu System - Miscellaneous Menu
No
If system is setup accordingly access level 3
code must be entered to access this menu.
This entry specifies if the analog outputs and
the concentration alarms status are updated
during calibrations or not:
Selecting Yes means during calibration
the analog outputs hold the last value
measured before calibration was started
concentration alarms possibly caused by
calibration gas concentrations are sup-
pressed
The alternative option (No) results in an ana-
log output following the measured value during
calibration. Alarms will be activated when the
currently measured values exceed the given
limits during calibration.
Note!
This behaviour may cause trouble when the
instrument is connected to e.g. a data aqui-
sition system.