Emerson Rosemount Xstream Series Instruction Manual page 246

Gas analyzer
Hide thumbs Also See for Rosemount Xstream Series:
Table of Contents

Advertisement

X-STREAM
Effect
Fluctuation or invalid
readout
(continued)
8-10
8-3
Solving Problems Not Indicated by Status Messages
Potential Cause
6. IR channel:
Source not connected or
defective
7. Analog preamplifier of affected
channel defective
8. Gas path(s) polluted
Instruction Manual
How to Solve
6. Check connections:
X3 (1/2) / source channel 1
X3 (4/5) / source channel 2
If source housing is cold:
Exchange both source in case
of dual channel analyzer /
replace source if need be
(see service manual)
7. Check measuring point
(
8-4-3-1-6, page 8-16)
8. Check analysis cells and
windows for pollution
Clean polluted parts (see
service manual)
Check gas paths for pollution
and clean gas paths if need be
9. Set ambient pressure to proper
value (
5-4-3-3, page 5-26)
Sensor failure (
message "PressSensor",
page 8-4)
10. Check temperature of gas
path(s)
Remove all sources of
condensation
Keep all temperatures at least
10 °C above sample gas
temperature
11. Replace BKS
Emerson Process Management GmbH & Co. OHG
HASAxE-IM-HS
05/2006