Connecting the power cable
The power cable is connected by screw termi-
nals, located at the analyzer's rear panel.
Supported wire cross sections:
Cable skinning length:
Hole diameter:
Screw thread:
Tightening torque, min:
To connect the terminals remove the cover at
the analyzer's rear panel (4 screws). Feed the
power cable through the lower edge protection,
through the lower strain-relief and around the
separation between power terminals and next
terminals row.
4-18
4-4-1 Installation - X-STREAM GP, X-STREAM GPS
ELECTRICAL SHOCK HAZARD!
Verify cables are disconnected from power
prior to working at the power terminals!
Use a power cable with plug for wall outlet so-
cket.
24 to 14 AWG (0.2 to 2.5 mm
no need to use wire end sleeves
0.354 inch (9 mm)
0.05 inch (1.2 mm)
M 2.5
3.5 in.lb (0.4 Nm)
The 3 rightmost terminals (between the fuse
holders) are reserved for power.
Fig. 4-16: Power terminals
Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASAxE-IM-HS
07/2006
2
),
Line (L)
Protective Earth (PE)
Neutral (N)