Connecting The Waste Line - Thermo Scientific Vanquish VF-D20 Operating Manual

Charged aerosol detectors
Hide thumbs Also See for Vanquish VF-D20:
Table of Contents

Advertisement

5.6.4

Connecting the Waste Line

Parts required
Tools required
Preparations
Follow these steps
Vanquish Charged Aerosol Detectors Operating Manual
NOTICE Extending or combining the waste line can cause a backpressure
in the line that can lead to an excess liquid build-up inside the detector.
Avoid extending or combining the waste line. Make sure that the waste
line drains freely into waste.
Waste line for connection between waste outlet and waste container
Tubing cutter (optional)
1. Route the waste line through the detector waste outlet in the
system base to waste. Follow the instructions in the Vanquish
System Operating Manual.
2. On the waste outlet on the detector, remove the plug.
1. Route the waste line from the system base to the detector, through
the tubing guides of the system modules below the detector.
2. On the waste outlet on the detector, connect the waste line.
Figure 22: Waste line connected to the waste port
No.
Description
1
Waste outlet
2
Waste line
5
Installation