Vanquish Charged Aerosol Detectors Operating Manual
4. Grasp the detector by the carrying handles. Slowly and carefully, lift
the detector out of the shipping container.
If the foam spacer under the detector is lifted out of the shipping
container with the detector, remove it from the detector before you
continue.
Figure 5: Carrying handles on the detector
No.
Component
1
Carrying handles
2
Attachment screw (one on each carrying handle)
5. Place the detector on a stable surface.
6. Transport the detector by the carrying handles to the installation
site, if it is not already there, and place it in the system stack (see
the System Arrangement section in this operating manual).
7. On each carrying handle, loosen the attachment screw until the
carrying handle is moveable in the rail. Do not remove the screws
from the carrying handles completely.
8. Slide off the carrying handles from the rails towards the rear of the
detector.
Figure 6: Sliding off the carrying handle from the left rail
TIP Keep the shipping container, the carrying handles with the
attachment screws, and all packing material. These items will be
needed if the detector is transported to a new location or shipped.
4
Unpacking and Transport
Page 41