Thermo Scientific Vanquish VF-D20 Operating Manual page 102

Charged aerosol detectors
Hide thumbs Also See for Vanquish VF-D20:
Table of Contents

Advertisement

6
Operation
Page 102
Parameters
Evaporator temperature
control
Evaporator
temperature nominal
Evaporator
temperature
Power function
Leak detection
Vanquish Charged Aerosol Detectors Operating Manual
Description
Temperature control for the evaporation tube is enabled
as a standard when the detector is shipped.
Specifies the setpoint for the evaporation temperature.
As a standard, the Evaporator Temperature Nominal is set
to 35 °C.
For further information, see
Temperature, page
107.
Displays the actual temperature of the evaporation tube.
The actual temperature depends on several factors,
including mobile phase composition and temperature,
pump flow rate, gas temperature, and ambient
temperature.
As a standard, the temperature unit is °C.
Sets by what factor a linearization is applied to the data
output. As a standard, the Power Function is set to 1.00.
For further information, see
Value, page
109.
Leak detection is enabled as a standard when the detector
is shipped (Leak Sensor Mode = Enabled). This is the
preferred setting.
section 6.7.5 Evaporation
section 6.7.7 Power Function