Thermo Scientific Vanquish VF-D20 Operating Manual page 64

Charged aerosol detectors
Hide thumbs Also See for Vanquish VF-D20:
Table of Contents

Advertisement

5
Installation
Follow these steps
Page 64
1. On the front of the detector, remove the dust seal plug from the
opening for the nebulizer.
Keep the dust seal plug to have it available when transporting the
detector.
Figure 13: Dust seal plug installed in the nebulizer opening
2. Insert the nebulizer in the opening in the front panel of the detector.
Observe the following:
TIP To facilitate inserting the nebulizer in the nebulizer opening,
moisten the sealing ring on the sealing bushing with deionized
water.
The nebulizer tip must point to the opening in the middle of the
nebulizer opening. Avoid the nebulizer tip from touching the
sides of the nebulizer opening, as this can damage the tip.
Vanquish Charged Aerosol Detectors Operating Manual