Before Starting Sample Analysis
Vanquish Charged Aerosol Detectors Operating Manual
Before starting an analysis:
Make sure that the pump flow rate that is delivered to the charged
aerosol detector is within the specified range. See
page
164.
To deliver pump flow to the charged aerosol detector, follow these
steps:
a) Turn on the detector gas flow. See
pump flow must be turned off. Wait at least 5 minutes before
you continue.
b) Turn on the pump flow to the detector. Have the pump deliver
flow to the detector.
Check the liquid level in the solvent reservoirs. Verify that the
amount of solvent is sufficient for the analysis.
Verify that the doors of the modules in the Vanquish system are
closed.
Make sure that the chromatography system is properly equilibrated.
System equilibration should include the following operations:
Purging the pump (all channels)
Flushing the entire chromatography system with the starting
solvent to rinse out any solvent from a previous analysis run
Warming up (or cooling down) all temperature-controlled
devices in the system to the starting temperature
Temperature-controlled devices can be, for example,
Column compartment and post-column cooler
Thermostatting in the autosampler
Flow cell in a fluorescence detector
Evaporation tube in a charged aerosol detector
Turning on the lamp (or lamps) in the UV/VIS detector
Monitoring the pump pressure and pressure ripple and checking
that the pressure is stable and the ripple within reasonable
limits for the application
6
Operation
section 9.1,
section 6.5.3, page
99. The