Installation
5.6.4

Installing the In-Line Filter (optional)

Parts required
Follow these steps
Page 90
The in-line filter can be installed in the flow path before the detector
optionally to prevent particulate matter from entering the nebulizer.
In-line filter, including pre-installed filter frit (from detector ship kit)
Figure 13: In-line filter
1. Locate and unpack the in-line filter.
2. Open the filter cap of the filter assembly.
3. Rinse the filter housing with deionized water.
4. Install the filter cap, and tighten it carefully until contact between
the cap and filter is felt. The connection should be finger-tight.
NOTICE Do not use a wrench or pliers to close the filter cap. Do not
overtighten the filter cap, as this can crush the filter.
AVIS N'utilisez pas de clé à molette ou des pinces pour fermer le
capot du filtre. Ne serrez pas de manière excessive le capot du filtre,
cela peut casser le filtre.
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
No.
Description
1
Filter cap (end nut)
2
Filter housing