Thermo Scientific Vanquish VF-D20 Operating Manual page 147

Charged aerosol detectors
Hide thumbs Also See for Vanquish VF-D20:
Table of Contents

Advertisement

Vanquish Charged Aerosol Detectors Operating Manual
4. Remove the slide-in module from the enclosure.
The following steps require a team effort:
a) Take the slide-in module on both sides from below.
b) Pull the slide-in module from the rails towards the front.
c) Place the slide-in module on a clean and stable surface.
5. Return the slide-in module:
a) To request a dedicated packaging for the slide-in module and for
the appropriate procedure for returning the module, contact
your local Thermo Fisher Scientific support organization.
b) Pack the slide-in module in the dedicated packaging.
NOTICE The packaging for the slide-in module differs from the
packaging for the complete module. Shipping the slide-in module or
the detector incorrectly leads to damage on the device. Always pack
and ship the slide-in module and the detector in the respective,
dedicated packaging.
CAUTION—Possible Contamination
Hazardous substances may have contaminated the device during
operation and may cause personal injury to service personnel.
Decontaminate all parts of the device that you want to return for
repair.
Fill in and sign the Decontamination Certificate, which is part of the
Service Return Form. Sign the certificate to confirm that the device
has been properly decontaminated and that it is free of hazardous
substances.
Thermo Fisher Scientific refuses to accept devices for repair if the
Decontamination Certificate is missing.
7
Maintenance and Service
Page 147