Thermo Scientific Vanquish VF-D20 Operating Manual page 88

Charged aerosol detectors
Hide thumbs Also See for Vanquish VF-D20:
Table of Contents

Advertisement

6
Operation
MUTE ALARM
GAS
When the detector is connected in the Chromeleon software
Page 88
If beeping alerts you that the detector firmware detected a problem, for
example leakage, beeping continues until you turn off the beep. Pressing
this button turns off the beep for the current alarm. As a standard,
beeping starts again after 10 minutes if the problem is still present or if
the firmware detected another problem.
Pushing the Gas button will perform the following actions:
Turn the detector gas flow on and off
Turn the drain pump on and off
After you pushed the button for turning off, the drain pump will
remain enabled for a short period longer to remove any remaining
mobile phase.
The LED next to the button indicates the gas status:
LED
Description
Off (dark)
Green
The button functionality is as follows when the detector is connected in
the Chromeleon software:
No injection (sample) or sequence or baseline monitoring is running:
All functions are available from the keypad.
An injection (sample) or sequence or baseline monitoring is running:
The Mute Alarm function remains available from the keypad,
allowing you to turn off the beep for the current alarm.
Vanquish Charged Aerosol Detectors Operating Manual
The detector gas flow and drain pump are turned off.
The detector gas flow and drain pump are turned on.