Thermo Scientific Vanquish VF-D20 Operating Manual page 122

Charged aerosol detectors
Hide thumbs Also See for Vanquish VF-D20:
Table of Contents

Advertisement

7
Maintenance and Service
Parts required
Preparations
Page 122
Suitable cleaning detergent (or disinfectant)
Purified water
Lint-free cloths or wipes
CAUTION—Explosive Gas Mixtures from Alcoholic Cleaning Detergents
Alcohol-containing cleaning detergents may form flammable and
explosive gas mixtures when exposed to air.
Use such cleaning detergents only when required and only in
adequately ventilated rooms.
Avoid open flames or exposure to excessive heat during the cleaning
process.
Wipe the cleaned components thoroughly dry after cleaning. Do not
operate the detector before it is completely dry.
NOTICE Observe the following:
Only use cleaning detergents that will not damage the surfaces of
the system.
Never use sharp tools or brushes for cleaning any surfaces.
Do not use sprays for cleaning.
Prevent cleaning detergent from entering the flow path.
Do not use excessively wetted cloth or wipes for cleaning. Prevent
any liquids from entering the functional components of the detector.
Liquids can cause a short circuit when getting in contact with the
electronic components.
Turn off the power to the detector and disconnect the power cord from
the power source.
Vanquish Charged Aerosol Detectors Operating Manual