Vanquish Charged Aerosol Detectors Operating Manual
3. Push the protective cover upward to release it from the two bottom
screws (no. 1 in
Figure
front panel (no. 2 in
Figure 31: Removing the protective cover
The gas filters are uncovered now:
Figure 32: Gas filter assembly
No.
Description
1
Charcoal gas filter
2
Elbow fittings
4. Disconnect the gas filter assembly from the gas connectors on the
detector:
a) On the two (push-in) elbow fittings on the detector gas
connectors, push in the locking ring on each elbow fitting with
the multitool (no. 1 in
fittings connected to the gas filter assembly.
7
31). Lift the protective cover away from the
Figure
31).
No.
Description
3
HEPA gas filter
4
Gas filter clamps
Figure 33