Thermo Scientific Vanquish VF-D20 Operating Manual page 79

Charged aerosol detectors
Hide thumbs Also See for Vanquish VF-D20:
Table of Contents

Advertisement

Vanquish Charged Aerosol Detectors Operating Manual
3. On the leak tray of the charged aerosol detector, route the capillary
through the capillary clip and secure it behind the clip.
a) Pull the capillary clip carefully to the right to open the clip.
b) Position the capillary behind the capillary clip.
c) Carefully release the clip and take care not to clamp the
capillary. The capillary must be secured behind the clip.
Figure 24: Securing the capillary behind the capillary clip on the leak tray
No.
Description
1
Capillary from the flow cell
4. Connect the inlet capillary to the nebulizer.
Figure 25: Series flow connection between diode array detector and
charged aerosol detector (example)
5
No.
Description
2
Capillary clip
Installation
Page 79