7
Maintenance and Service
Follow these steps
Page 146
1. Loosen the four captive screws on the front left and front right of
the detector.
Figure 38: Captive screws on the slide-in module
2. Push all tubing and capillaries, which are present in the tubing chase
of the Vanquish system modules, into the tubing chase. Otherwise,
you will not be able to remove the slide-in module properly from the
enclosure in the next step.
3. Grasp the slide-in module by the leak tray, and pull the module out
of the enclosure by approximately 10 cm.
Figure 39: Pulling out the slide-in module
NOTICE The slide-in module can fall down when pulling it out of the
enclosure too far. Pull out the slide-in module just far enough so
that you can grasp it on both sides from below.