Advertisement

5
Installation
Page 54
The use of a non-bottled gas source is recommended, such as
A nitrogen generator with suitable air compressor or with a
suitable in-house air source
Medical-grade nitrogen from an in-house liquid nitrogen source
TIP Use of smaller compressed nitrogen cylinders or liquid nitrogen
containers may be a quick and easy source of nitrogen during
installation. However, it is not recommended for routine operation.
The gas consumption of the detector is typically ≤ 4 L/min.
The supply gas must be free from water vapor, free from particles,
and must not contain non-volatile hydrocarbons, such as
compressor oils.
If necessary, use a sub-micron particle filter or carbon filter in
combination with a water condensation trap located close to the gas
source.
Make sure that the gas supply is ready for operation. Follow the
instructions in the User Documentation for the gas supply.
Vanquish Charged Aerosol Detectors Operating Manual