Follow these steps
Vanquish Charged Aerosol Detectors Operating Manual
1. Start the Chromeleon 7 Instrument Configuration Manager or the
Chromeleon 6.8 Server Configuration program.
2. Perform a firmware update from the General tab page in the
configuration dialog box for the detector. For details, refer to the
Chromeleon Help.
NOTICE A firmware downgrade or incomplete firmware update may
result in loss of functionality or malfunctioning of the detector.
Do not interrupt communication between the Chromeleon software
and the detector at any time during the procedure.
At the beginning of the update process, a message appears showing
the firmware version currently installed in the detector and the
version that will be transferred from the Chromeleon software. If the
firmware installed in the detector is a later version than the version
in Chromeleon, cancel the download.
The firmware update may take several minutes.
Monitor the Audit Trail of the Chromeleon Instrument Configuration
Manager (or Server Configuration program) to see whether the
firmware update was successful or failed.
If the firmware update failed, turn the device off and on again and
repeat the firmware update.
If the firmware update fails repeatedly, contact Thermo Fisher
Scientific Technical Support for assistance.
3. After a successful firmware update, requalification of the detector
may be required. See the release notes for a recommendation.