PrimeLab 1.0 MULTITEST User Manual page 97

Photometer
Table of Contents

Advertisement

COD MR
(80)
Continued...
12) The determined result is immediately displayed. Press the "OK"
4
button to save the result together with date and time.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
Note(s)
->
Suspended particles in the zero cell and / or the sample cell lead to wrong
test results. Make sure that any existing suspended solids have settled to the
bottom of the cell and are not disturbed by the insertion into the PrimeLab.
->
Both cells used for the measurement (zero / sample) must be from the same
production batch. The cell used for ZERO can be kept for other tests (of the
same batch) but must be stored in the dark.
->
For COD content above 1500 mg/l, the use of another method (COD HR) is
recommended to achieve accurate results.
->
Never insert hot cells into the PrimeLab measuring chamber!
->
This method is not suitable for water samples with Chloride values higher
than 1,000 mg/l.
64
0 - 1500 mg/l
Name on device: 80-COD-MR-pre
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
(O )
16 mm cuvette
2