PrimeLab 1.0 MULTITEST User Manual page 105

Photometer
Table of Contents

Advertisement

Copper
(18)
(free/combined/total)
Continued...
Press button 4 (OK) to continue the measurement to determine the total copper
!
content (tCu) and later also the combined copper (cCu). If this is not desired the
measurement procedure can be terminated by pressing "HOME". In this case only
8) Remove the lid again and add a "Copper N° 2 Photometer" tablet into
the same sample you have just used for determining "fCu" (free copper).
9) Crush the tablet with a clean stirrer until it has completely dissolved.
10) Replace the lid on the cuvette, swirl it carefully and insert it into the
sampling chamber of the PrimeLab. Align the arrow on the front of the
4
cuvette.
Press TEST.
11) The total result is displayed immediately, divided in
"fCu" = "free copper"
2
"cCu" = "combined copper"
3
"tCu" = "total copper"
To scroll through the values please used the arrow up and down buttons.
4
The result is saved when button 4 is pressed.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
72
0.05 - 5 mg/l
Name on device: 18-Copper-tab
the valule for free copper (fCu) is stored.
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
(Cu)
Tablet
Continued...
72