PrimeLab 1.0 MULTITEST User Manual page 195

Photometer
Table of Contents

Advertisement

Nitrate
(34)
Continued...
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
Note(s)
->
(*) Conversion of the measurement result to NO .
->
Make sure that the sample contains no air bubbles in the measurement.
If this is the case, remove the air bubbles by tapping it with the cuvette.
->
Extend the listed measurement range to 0 - 110 mg/l (N)
by taking 1 ml of sample water plus 9 ml of distilled water.
The test result displayed on the screen needs to be multiplied by 10.
->
Best results are achieved between 0 – 6 mg/l (N) / 0 – 25 mg/l (NO ).
If your water sample is likely to contain more Nitrate, dilute the sample to
bring it into the above mentioned measurement range. You can do so, by
(e.g.) diluting 5 ml sample water plus 5 ml deionized water, measure as
usual, multiply result x 2.
162
0.0 - 11.0 mg/l
Name on device: 34-Nitrate-pow
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
(N)
3
Liquid
+Powder
3
162

Hide quick links: