Hardness-Calcium
(78)
Continued...
10) Press button 4 to save the result together with date and time in the
4
device.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
Note(s)
->
(*) Conversion to °dH, °eH and °fH
->
If your reading is towards the upper limit of the test a dilution is
recommnded.
->
If your sample water is very alkaline or acidic it should be brought within the
pH range between 4 and 10 by adding 1 mol/l acetic acid or resp. 1 mol/l
caustic soda.
->
Ensure that you are using exactly 10 ml sample water volume.
->
Disturbance values in this measurement value:
Magnesium > 200 mg/l CaCO3
Zinc > 5 mg/l
Iron > 10 mg/l.
94
0 - 500 mg/l
Name on device: 78-Hard-Cal-tab
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
(CaCO )
Tablet
3
94