Menu: SET (Setup)
Continued...
Calibration
The following refers to calibration of ID 111/112/113 (PTSA/Turbidity/Fluorescein):
As this Parameter uses an indirect light from above, it is essential to always ensure the
correct amount of water in the vial, which is why exactly 10 ml of liquid should be taken
when doing a test. To achieve this, use the pipette which is part of the Adapter-kit. Please
change or clean the tip of the pipette after each measurement/calibration using destilled
water.
Please perform a new calibration for the parameters of the ID111 / 112 / 113 if the
measurement results obtained appear to be inaccurate or no longer correspond to
the value of the standard solutions supplied.
If you select „Turbidity-adapter":
You first have to perform a „PrimeLab calibration" !!! See page „SET 4"
Calibration steps, as shown below, will be displayd on the PrimeLab as well. Please
make sure the standard vials (0.5/10/1000 NTU) are free from fingerprints and dirt.
Clean vials properly by using the cloth provided in your kit, prior inserting into the
PrimeLab.
0.5 NTU standard vial: Let vial rest for at least 3 hours before use for the calibration
purposes. Turn vial TWO TIMES before you open it. Do not shake!
10 NTU standard vial: Turn vial THREE TIMES before you open it. Do not shake!
1000 NTU standard vial: Shake vial heavily before you open it!
Open „0.5 NTU" standard vial and insert without lid and with forward arrow into the
PrimeLab.
Set Turbidity adapter on top of the open cell.
Press TEST.
4
Remove vial from PrimeLab and close it properly for future use.
Open „10 NTU" standard vial and insert without lid and with forward arrow into the
PrimeLab.
Set Turbidity adapter on top of the open cell.
Press TEST.
4
Remove vial from PrimeLab and close it properly for future use.
Open „1000 NTU" standard vial and insert without lid and with forward arrow into
the PrimeLab.
Set Turbidity adapter on top of the open cell.
Press TEST.
4
SET 5
Use caution to shaking-/rest-instructions
on the standard vials (0.5/10/1000 NTU):
ID 112 is now calibrated successfully.