PrimeLab 1.0 MULTITEST User Manual page 201

Photometer
Table of Contents

Advertisement

Nitrite (HR)
(97)
Continued...
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
Notes
->
Filter sample if neccessary to test a clear sample.
->
Make sure the temperature of your sample does not exceed 30°C.
->
High chlorine levels (> 30 mg/l) interfere with this test and can influence the
reading.
->
The test needs to be performed without a delay. Place the vial into the
PrimeLab right after reagents have dissolved and lid is closed. Immediately
press TEST. It is essential for the accuracy of this test to keep the countdown
of 02:00 minutes right after dissolving the tablets/closing the lid/placing the
vial into the PrimeLab.
->
DO NEVER SHAKE THE VIAL!
->
For expected values below 400 mg/l it is strongly recommended - to achieve
the most accurate result - to better use ID 36 (Nitrite with powder reagents
0 - 200 mg/l; extended range 0 - 400 mg/l by 1:1 dilution).
168
0 - 1500 mg/l
Name on device: 97-Nitrite-HR-tab
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
(NaNO )
2
Tablet