PrimeLab 1.0 MULTITEST User Manual page 172

Photometer
Table of Contents

Advertisement

Legionella
(147)
Continued...
63) Transfer 1ml from LG-MHCB2 (right) to a new, clean small vial
„LG-CB". USE A CLEAN PIPETTE.
64) Place filled 1ml LG-CB vial into the adapter. Place the light shield
4
(PLSp-LS-1) on top and press TEST.
65) The determined result is immediately displayed. Press the "OK" button
4
to save the result together with date and time.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
B) Measurement procedure for B) Countdown + TEST
1) Eluate filtered particles by shaking for 2 minutes.
2) Press OK to start 02:00 minutes countdown.
4
3) Press OK to start 15:00 minutes countdown.
4
4) Press OK to start 05:00 minutes countdown.
4
5) Press OK to start 03:00 minutes countdown.
4
6) Press OK to start 10:00 minutes countdown.
4
7) Press OK to start 03:00 minutes countdown.
4
8) Press OK to start 03:00 minutes countdown.
4
9) Press OK to start 03:00 minutes countdown.
4
10) Press OK to start 03:00 minutes countdown.
4
11) Press OK to start 02:00 minutes countdown.
4
12) Press OK to start 05:00 minutes countdown.
4
13) Place filled 1ml LG-CB vial into the adapter. Place the light shield
(PLSp-LS-1) on top and press ZERO.
4
14) Place filled 1ml LG-CB vial into the adapter. Place the light shield
(PLSp-LS-1) on top and press TEST.
4
139
60 - 1000000 cfu/test
Name on device: 147-Legionella-liq
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
Liquid
Continued...
139