Chlorine Dioxide
(130)
Continued...
Note(s)
->
Alkalinity of more than 250 mg/l CaCO affect the measurement and the
color development. Neutralize to pH 6-7 with 1 N Hydrochloric Acid before
performing the test.
->
Acidity of more than 150 mg/l CaCO affect the measurement and the color
development. Neutralize to pH 6-7 with 1 N Sodium Hydroxide before
performing the test.
->
Bromine at all levels, Ozone (if >1.5 mg/l) and chlorine (if >6 mg/l) as well as
iodine will interfere with the reagent reaction and will be part of the displayed
value. Peroxides may interfere as well.
->
Organic chloramines may interfere. Monochloramines lead to higher
readings (~0.1 mg/l at 3 mg/l monochloramines after 1 minute).
->
If oxidized manganese or oxidized chromium is present in the sample,
sample needs to be pre-treated.
->
Hardness levels above 1000 mg/l CaCO affect the measurement
->
Extreme pH-values of the sample need to be corrected to pH 6-7 before
measuring the sample.
56
0.00 - 5.00 mg/l
Name on device: 130-Chl-Diox-pp
(ClO )
2
3
3
3
Liquid
+Powder