PrimeLab 1.0 MULTITEST User Manual page 191

Photometer
Table of Contents

Advertisement

Nickel (HR)
(90)
Continued...
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
Notes
->
Ferrous Iron will interfere with this test and can influence the readings.
->
High EDTA levels (> 25 mg/l) will interfere with this test and can influence
the reading (low reading).
->
High cobalt levels (> 0.5 mg/l) will interfere with this test and can influence
the reading (high reading).
->
Polyphosphates in the sample does not influence the reading.
158
0 - 7 mg/l
Name on device: 90-Nickel-HR-tab
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
(Ni)
Tablet
158