PrimeLab 1.0 MULTITEST User Manual page 266

Photometer
Table of Contents

Advertisement

Sodium Hypochlorite
(51)
Continued...
11) Add an "Acidifying GP" tablet to the water sample in the same
cuvette.
12) Crush the tablet with a clean stirrer until this is completely dissolved.
13) Replace the lid on the cuvette, swirl it carefully and insert it into the
sampling chamber of the PrimeLab. Align the arrow on the front of the
4
cuvette. Press TEST.
14) After the lapse of a 10 second countdown the determined result is
4
displayed. Press button 4 to save the test result together with date and
time on the device.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
Note(s)
->
The precision of the test results depends upon the precision of the diluting
procedure.
233
0.2 - 40 %
Name on device: 51-Sodium-Hypo-tab
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
(NaOCl)
Tablet
233

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents