PrimeLab 1.0 MULTITEST User Manual page 164

Photometer
Table of Contents

Advertisement

Iron in Oil
(149)
Adapter
Syringe
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5". Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
!
The individual steps of the measurement procedure are shown on the display.
1) Open the lid of a clean (new) beaker.
2) Add 3ml of TM-reagent-E into the beaker.
3) Add 1ml of TM-reagent-C into the beaker.
4) Add 5ml of TM-reagent-I into the beaker.
5) Close the lid of the beaker and shake/swirl beaker until reagents are
mixed homogenously.
6) Place 3ml-vial-adapter into the PrimeLab.
7) Fill a clean luer-lock syringe with exactly 3ml of the water-part of your
sample.
8) Empty syringe into a clean/new 3ml vial.
9) Place 3ml vial into the adapter and apply light shield.
One-Time-Zero:
Measure new ZERO
Use last ZERO
1
2
3
!
10) Press ZERO.
4
11) Remove light protection shield and 3ml vial.
12) Open the lid of a clean (new) beaker
131
50 - 500 mg/l
Name on device: 149-Iron-Oil-liq
(*)
(*)
Filter paper Filter adapter
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
-> Use last ZERO
4
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
2+
(Fe )
(*)
beaker
= continue with step 10)
= continue with step 11)
Liquid
TM-reagent-S
(TM149reagS)
TM-reagent-E
(TM149reagE)
TM-reagent-I
(TM149reagI)
TM-reagent-C
(TM149reagC)
Continued...
131