PrimeLab 1.0 MULTITEST User Manual page 86

Photometer
Table of Contents

Advertisement

Chlorine Dioxide
(64)
Continued...
8) Replace the lid on the cuvette, swirl it carefully and insert it into the
sampling chamber of the PrimeLab. Align the arrow on the front of the
4
cuvette. Press TEST.
9) The determined result is immediately displayed. Press the "OK" button
to save the result together with date and time.
4
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
B) Measurement procedure for "Chlorine dioxide without chlorine"
1) Fill 10ml of the sample water into a clean 24mm cuvette.
One-Time-Zero:
Measure new ZERO
Use last ZERO
1
2
3
!
2) Replace the screw lid on the cuvette and insert it into the PrimeLab
sampling chamber. Align the arrow on the front of the cuvette. Press
4
ZERO.
3) Remove the cuvette from the chamber and unscrew the lid.
4) Empty the cuvette except for a few drops. Then add 3 drops of "PL
DPD 1 A" and 3 drops of "PL DPD 1 B" to the remaining test liquid in the
cuvette.
5) Fill sample water into the same cuvette up to the 10ml level.
6) Replace the lid on the cuvette, swirl it carefully and insert it into the
sampling chamber of the PrimeLab. Align the arrow on the front of the
4
cuvette. Press TEST.
7) The determined result is immediately displayed. Press the "OK" button
4
to save the result together with date and time.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
53
0 - 15 mg/l
Name on device: 64-Chlorin-Dio-liq
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
-> Use last ZERO
4
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
(ClO )
2
= continue with step 2)
= continue with step 4)
Liquid
Continued...
53

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents