PrimeLab 1.0 MULTITEST User Manual page 140

Photometer
Table of Contents

Advertisement

Hydrocarbons
(160)
10 ml
24 mm
PLSp-ADP-TRB
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5". Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
!
The individual steps of the measurement procedure are shown on the display.
1) Fill a cleaned cell (24mm) with 2.5ml of your sample and 7.5ml of
deionised water
2) Close the lid. Shake the cell gently to mix the liquid.
3) Open the lid.
4.) Place the cell into the PrimeLab (use caution to the arrow on the cell).
Do NOT close the lid.
5.) Place TURBIDITY-ADAPTER on top of the open cuvette.
6.) Press TEST.
4
!
Note(s)
->
Interferences:
- Hazy test sample
- The presence of excessive detergents
107
Name on device: 160-Hydrocarbons
PLSp-PIP10ml
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
Interpretation:
0 = PASSED
OR = FAILED
0 - 1
PLSp-PIP10ml-tips
(YES / NO)
107