PrimeLab 1.0 MULTITEST User Manual page 100

Photometer
Table of Contents

Advertisement

Colour
(107)
(*)
10 ml
24 mm
Syringe
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5". Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
!
The individual steps of the measurement procedure are shown on the display.
Select either:
A) „True colour"
B) „Apparent colour"
A) Measurement procedure for „True colour"
1.) Unscrew the two halves of the filter holder and place one filter circle
(0.45µ) onto the base section.
2.) Screw the two parts together again, ensuring the O ring is correctly
located.
3.) Fill a clean luer-lock syringe (20 ml) with deionised water.
4.) Connect the syringe to the filter holder and discharge the syringe to
waste.
5.) Disconnect the syringe from the filter holder and refill syringe with
destilled water.
6.) Fill a clean luer-lock syringe (20 ml) with deionised water.
7.) Connect the syringe to the filter holder and discharge the syringe to
waste.
8.) Disconnect the syringe from the filter holder and refill syringe with
destilled water.
9.) Fill a clean luer-lock syringe (20 ml) with deionised water.
10.) Connect the syringe to the filter holder and discharge the syringe to
waste.
11.) Disconnect the syringe from the filter holder and refill syringe with
destilled water.
67
0 - 500 mg/l
Name on device: 107-Colour
(*)
(*)
Filter paper
Filter adapter
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
(Pt-Co)
Continued...
67