PrimeLab 1.0 MULTITEST User Manual page 264

Photometer
Table of Contents

Advertisement

Silica (HR)
(50)
Continued...
11) After the lapse of a 2 minute countdown the determined result is
displayed. Press button 3 to convert this into various measurement units
3
(*).
12) Press button 4 to save the result together with date and time in the
4
device.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
Note(s)
->
(*) Conversion to Si
->
The temperature of the water sample must be between 15°C and 25°C to
ensure precise measurements.
->
Sulfide in the water sample will influence the measurement result.
->
Larger amounts of iron falsify the measurement result.
->
Phosphate content in the water higher that 60 mg/l will falsify the
measurement result.
231
0 - 100 mg/l
Name on device: 50-Silica-HR-pow
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
Powder
(SiO )
2
231