PrimeLab 1.0 MULTITEST User Manual page 101

Photometer
Table of Contents

Advertisement

Colour
(107)
Continued...
12.) Fill a clean luer-lock syringe (20 ml) with deionised water.
13.) Connect the syringe to the filter holder and discharge the syringe to
waste.
14.) Disconnect the syringe from the filter holder and refill syringe with
destilled water.
15.) Fill a clean luer-lock syringe (20 ml) with deionised water.
16.) Connect the syringe to the filter holder and discharge the syringe to
waste, down to the 10 ml mark.
17.) Filter the remaining 10 ml of destilled water into a clean 24mm vial.
18.) Close the lid, place the cell into the photometer (use caution to the
arrow on the cell) and press ZERO.
4
19.) Remove the cell and open the lid.
20.) Disconnect the syringe from the filter holder and refill syringe with
SAMPLE water.
21.) Connect the syringe to the filter holder and discharge the syringe to
waste.
22.) Disconnect the syringe from the filter holder and refill syringe with
SAMPLE water.
23.) Connect the syringe to the filter holder and discharge the syringe to
waste.
24.) Disconnect the syringe from the filter holder and refill syringe with
SAMPLE water.
25.) Connect the syringe to the filter holder and discharge the syringe to
waste, down to the 10 ml mark.
26.) Filter the remaining 10 ml of SAMPLE water into a clean 24mm vial.
27.) Close the lid, place the cell into the photometer (use caution to the
4
arrow on the cell) and press TEST.
68
0 - 500 mg/l
(Pt-Co)
Name on device: 107-Colour
Continued...
68

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents