Chlorine
(11)
(free/combined/total)
Continued...
After pressing the button 4 (OK) the measurement is continued to determine the
!
total chlorine (tCl) and later also bonded chlorine (cCl). If this is not required the
measurement can be termianted by pressing "HOME". IN this case only the value
8) Remove the lid again and add a "DPD N° 3 Photometer" tablet into the
same sample you have just used for determining "free chlorine".
9) Crush the tablet with a clean stirrer until it has completely dissolved.
10) Replace the lid on the cuvette, swirl it carefully and insert it into the
sampling chamber of the PrimeLab. Align the arrow on the front of the
cuvette.
4
Press TEST.
11) After the lapse of a 2 minute countdown the total result is displayed,
divided in
2
3
"fCl" = "free chlorine"
"cCl" = "combined chlorine"
"tCl" = "total chlorine".
4
Scroll through the results using the buttons "arrow up" and "arrow down".
Press button 4 to save the result.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
37
0.00 - 8.00 mg/l
Name on device: 11-Chlorine-tab
for free chlorine (fCl) is stored.
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
(Cl )
Tablet
2
Continued...
37