Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
B) Measurement procedure for „Apparent colour"
1.) Fill a cleaned cell (24mm) with 10 ml deionised water.
One-Time-Zero:
Measure new ZERO
Use last ZERO
1
2
3
!
2.) Close the lid, place the cell into the photometer (use caution to the
4
arrow on the cell) and press ZERO.
3.) Remove the cell and open the lid.
4.) Empty the cell.
5.) Rinse cell with this water sample and fill up to 10 ml mark.
6.) Close the lid.
7.) Place the cell into the device. Again, please use caution to the arrow
4
on the cell. Than press TEST.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
69
0 - 500 mg/l
Name on device: 107-Colour
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
-> Use last ZERO
4
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
(Pt-Co)
= continue with step 2)
= continue with step 4)
Continued...