14.) Apply the stopper.
15.) Gently shake glass bottle for 01:00 minute.
16.) Press TEST to start 01:00 minute countdown.
4
17.) Remove the stopper.
18.) Add 10 drops "PL DissOx 3" into the glass bottle.
19.) Apply the stopper.
20.) Gently shake glass bottle for 01:00 minute.
21.) Press TEST to start 01:00 minute countdown.
4
22.) Transfer 10 ml into the same vial, used for ZERO.
23.) Place the cell into the device. Again, please use caution to the arrow
4
on the cell. Than press TEST.
Notes
->
Make sure the 50 ml glass bottle is really filled up to the top and the water
will run out when applying the stopper
87
0 - 10 mg/l
Name on device: 163-DissOxyg-liq