PrimeLab 1.0 MULTITEST User Manual page 62

Photometer
Table of Contents

Advertisement

Bromine
(128)
Continued...
Note(s)
->
Alkalinity of more than 250 mg/l CaCO affect the measurement and the
color development. Neutralize to pH 6-7 with 1 N Hydrochloric Acid before
performing the test.
->
Acidity of more than 150 mg/l CaCO affect the measurement and the color
development. Neutralize to pH 6-7 with 1 N Sodium Hydroxide before
performing the test.
->
Other oxidants, such as chlorine, chlorine dioxide, ozone, peroxides as well
as iodine will interfere with the reagent reaction and will be part of the
displayed value.
->
If oxidized manganese or oxidized chromium is present in the sample,
sample needs to be pre-treated.
->
Hardness levels above 1000 mg/l CaCO affect the measurement
->
Extreme pH-values of the sample need to be corrected to pH 6-7 before
measuring the sample.
29
0.0 - 4.5 mg/l
Name on device: 128-Bromine-pp
(Br )
2
3
3
3
Powder
29