Chlorite
(106)
Continued...
11) Fill sample water into the same cuvette up to the 10ml level.
12) Add 3 drops "PL DPD 3 C" liquid reagent to the test liquid in the
cuvette.
13) Replace the lid on the cuvette, swivel it carefully and insert it in the
sampling chamber of the PrimeLab. Align the arrow on the front of the
4
cuvette. Press TEST.
14) Wait for the lapse of a 2 minute countdown.
15) Remove the cuvette from the chamber and unscrew the lid.
16) Add 3 drops "PL DPD Acidifying" liquid reagent to the test liquid in the
cuvette.
17) Replace the lid on the cuvette, swivel it carefully and insert it in the
sampling chamber of the PrimeLab. Align the arrow on the front of the
4
cuvette. Press TEST.
18) Wait for the lapse of a 2 minute countdown.
19) Remove the cuvette from the chamber and unscrew the lid.
20) Add 3 drops "PL DPD Neutralising" liquid reagent to the test liquid in
the cuvette.
21) Replace the lid on the cuvette, swivel it carefully and insert it in the
sampling chamber of the PrimeLab. Align the arrow on the front of the
cuvette. Press TEST.
22) The determined result is immediately displayed. Press the "OK" button
4
to save the result together with date and time.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
Note(s)
->
DPD 1 A and DPD 1 B reagent MUST be added to the vial BEFORE
water sample is added to avoid mis-readings!
->
The range of the side measurements performed during the Chlorite tests
may be up to 8 mg/l whilst it is rather unlikely that the range of Chlorite can
be tested up to 8 mg/l.
58
0 - 8 mg/l
Name on device: 106-Chlorite-liq
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / Underrange!":
(Cl )
2
Liquid
58