Iron (HR)
(30)
Continued...
8) Add 15 drops "PL Iron HR 1" liquid reagent to the test liquid in the
cuvette.
9) Replace the lid on the cuvette, swirl it carefully and insert it into the
sampling chamber of the PrimeLab. Align the arrow on the front of the
cuvette. Press TEST.
10) After the lapse of a 5 minute countdown the determined result is
displayed. Press button 4 to save this together with date and time in the
device.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
Note(s)
->
(*) Special accessories required / not included as standard equipment!
->
If the sample needs to be filtered (dissolved iron) it must be ensured that
after filtration there are no more undissolved iron parts in the sample. If you
are unsure, please repeat the filtration.
->
High nitrite values in the sample water can influence the measurement. If the
sample water turns to red or pink after adding "PL Iron HR 1" drops a new
sample needs to be taken with 0.1 g "TN1" powder added to it. After adding
the powder wait for 2 minutes and then start the measurement procedure as
described on the front page.
128
0 - 30 mg/l
Name on device: 30-Iron-HR-liq
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
2+/3+
(Fe
)
Liquid
128